Data Sheet

GPT-9500 Series User Manual
246
[:SOURce]:SAFEty:STEP<n>:DC:TIME:FALL
Sets the fall time value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP2:DC:TIME:FALL 0
Sets the fall time value for DCW of step2 to off.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:FALL?
Returns the fall time value for DCW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return Parameter: <NR3>
Example: SAFE:STEP2:DC:TIME:FALL?
> +0.000000E+00
The fall time value for DCW of step2 set is off.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel[:HIGH]
Sets the output channel status for DCW of scanning test high
voltage.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP2:DC:CHAN (@(1,3))
Sets the output channel for DCW of step2 to channel 1 and 3
HIGH output
Example: SAFE:STEP2:DC:CHAN (@(0))
Sets the output channel for DCW of step2 to off for all HIGH
output.