Data Sheet

REMOTE CONTROL
259
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel[:HIGH]
Sets the output channel status for OSC of scanning test high
voltage.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))
Example: SAFE:STEP4:OSC:CHAN (@(1))
Sets the output channel for OSC of step4 to channel 1 HIGH
output
Example: SAFE:STEP4:OSC:CHAN (@(0))
Sets the output channel for OSC of step4 off for all HIGH
output.
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel[:HIGH]?
Returns the output channel status for OSC of scanning test
high voltage.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: (@(CH))
Example: SAFE:STEP4:OSC:CHAN?
> (@(1))
The output channel for OSC of step4 is set to channel 1 HIGH
output.
[:SOURce]:SAFEty:STEP<n>:OSC:CHANnel:LOW
Sets the output channel status for OSC of scanning test
return(low).
Parameter<n>: <NR1>(1 ~ 99)
Parameter: (@(CH))