Data Sheet

OPERATION
43
Setting the MANU Test Mode
Background
Essentially, there are 5 modes, AC Withstand
(ACW), DC Withstand (DCW), Insulation
Resistance (IR), Pause (PA) and Open Short
Check (OSC) modes. Precisely, the previous 3
(ACW, DCW and IR) are for both MANU and
AUTO tests, whereas the rest 2 (PA and OSC)
are for AUTO test only.
Steps
1. Press the MANU/AUTO soft-
key to select MANU option.
2. Press the EDIT soft-key followed
by clicking the MODE soft-key.
3. Navigate the scroll wheel to
toggle between 5 modes. Further
press the SAVE soft-key to
confirm the selection.
ACW
DCW
IR
PA
OSC
AC Withstand (MANU, AUTO)
DC Withstand (MANU, AUTO)
Insulation Resistance (MANU, AUTO)
Pause action(AUTO)
Open Short Check action(AUTO)
Selected MANU Test Mode
4. Press the SAVE soft-key to complete.