Data Sheet

OPERATION
67
Setting OFFSET reference value
Background
The OFFSET is used to determine the offset
resistance of the tester. It is imperative to turn
ON GR MODE before setting OFFSET value.
When an OFFSET is performed, the reference is
automatically set to the measured resistance.
This function is only applicable to IR test.
Steps
1. Press the MANU/AUTO soft-key
to select MANU option.
2. Press the EDIT soft-key followed by
clicking the PAGE key three times.
x 3
3. Press the OFFSET soft-key followed
by using the scroll wheel to turn
ON/OFF the OFFSET function.
OFFSET
ON, OFF
Turn ON OFFSET
4. Press the SAVE soft-key to complete.