Data Sheet

OPERATION
99
AFTER FAIL
Background
The AFTER FAIL setting refers to the ensuing action
after FAIL judgment is shown on the display.
1. Use the PREV & NEXT keys or
scroll wheel to move to the AFTER
FAIL field.
2. Press the ENTER soft-key followed
by using the scroll wheel to select
an option.
CONT.
Next step keeps going even after
FAIL judgment appears.
STOP
AUTO test will be stopped
immediately after FAIL judgment of
one of the steps appears. Only the
STOP key can be pressed to return
to the READY status.
RESTART
AUTO test will be stopped
immediately after FAIL judgment of
one of the steps appears. User can
press START key to restart the
AUTO test from the 1
st
step.
3. Press the ENTER soft-key to
confirm the setting