Data Sheet

GPT-9500 Series User Manual
122
Running an Automatic Test
Background
An automatic test can be run when the tester is
in READY status.
Note
The tester cannot start to run an AUTO test
under the following conditions:
Any protection modes have been tripped.
The INTERLOCK function is ON and the
Interlock wire is not shorted in the Interlock
terminal (page 194).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START
button is pressed immediately after the STOP
button (<0.5s).
Warning
Do not touch any terminals, test leads or the DUT
when a test is running.
Steps
1. Ensure tester is in READY status for the AUTO
test to come. If it is under VIEW status, press
the STOP key to return to the READY status.
Normal AUTO
Display
READY status