Data Sheet

UTILITY
165
Double Action Setting
Description
Double Action function is a safety feature used
to prevent accidentally starting a test. Normally
to start a test, the START button is pressed
when the tester is in the READY status. To start
a test of either MANU or AUTO when Double
Action is ON, the STOP button must first be
pressed, followed by the START button within
500ms, during which the READY status is in
white color; otherwise the status of READY
remains in grey color indicating safety
protection is activated.
Steps
1. Press the UTILITY key on the front
panel followed by pressing the
PAGE DOWN soft-key repeatedly
until the Test page appears.
UTILITY
2. Use the scroll wheel to move to the
Double Action field.