Data Sheet

REMOTE CONTROL
233
[:SOURce]:SAFEty:RESult:STEP<n>:OMETerage?
Returns the OUTPUT METER readings of selected STEP.
Parameter<n>: <NR1>(1 ~ 99)
Return Parameter: <NR3>output meter
Example: SAFE:RES:STEP2:OMET?
> +5.000000E+02
The OUTPUT METER result of the second STEP is 0.500kV.
[:SOURce]:SAFEty:RESult:STEP<n>:MMETerage?
Returns the MEASURE METER readings of selected STEP.
Parameter<n>: <NR1>(1 ~ 99)
Return Parameter: <NR3>measure meter
Example: SAFE:RES:STEP2:MMET?
> +5.000000E-05
The MEASURE METER result of the second STEP is 0.05mA.
[:SOURce]:SAFEty:STEP<n>:DELete
Delete all setting value in selected Step.
Parameter<n>: <NR1>(1 ~ 99)
Example: SAFE:STEP1:DEL
Deleting the STEP1 setting value in the memory.
[:SOURce]:SAFEty:STEP<n>:SET?
Returns the all setting values in selected STEP.
Parameter<n>: <NR1>(1 ~ 99)
Example: SAFE:STEP1:SET?