[:SOURce]:SAFEty:STEP<n>:AC:REF
Sets the test offset value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: <NRf>
Example: SAFE:STEP1:AC:REF 0.00001
Sets the test offset value for ACW of step1 to 0.01mA.
[:SOURce]:SAFEty:STEP<n>:AC:REF?
Returns the test offset value for ACW in selected step.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: <NR3>
Example: SAFE:STEP1:AC:REF?
> +1.000000E-05
The test offset value for ACW of step1 is 0.01mA.
[:SOURce]:SAFEty:STEP<n>:AC:GROUndmode
Sets the ground mode for ACW in selected step to on or off.
Parameter<n>: <NR1>(1 ~ 99)
Parameter: 0 | 1 | OFF | ON
Example: SAFE:STEP1:AC:GROU 0
Sets the ground mode for ACW of step1 to off.
[:SOURce]:SAFEty:STEP<n>:AC:GROUndmode?
Returns the ground mode status for ACW.
Parameter<n>: <NR1>(1 ~ 99)
Return parameter: 0 | 1
Example: SAFE:STEP1:AC:GROU?