Data Sheet

OPERATION
89
FAIL Timing
Diagrams
The timing diagrams below show the ACW, DCW
and IR timing for the START status, TEST status
and FAIL judgment.
ACW FAIL Timing
DISCHARGE
Output V
time
START
TEST
FAIL
RAMP
UP
TEST TIME
DCW FAIL Timing
DISCHARGE
Output V
time
START
TEST
FAIL
RAMP
UP
TEST TIME
IR FAIL Timing
DISCHARGE
Output V
time
START
TEST
FAIL
RAMP
UP
TEST TIME