9
Test Voltage
Measurement Range / Accuracy
50V V<500V
0.1MΩ~1MΩ±(5% of reading + 3 count)
1 MΩ~50MΩ±(5% of reading + 1 count)
51MΩ~2GΩ±(10% of reading + 1 count)
500V V 1000V
0.1MΩ~1MΩ±(5% of reading + 3 count)
1 MΩ~500MΩ±(5% of reading + 1 count)
501MΩ~10GΩ±(10% of reading + 1 count)
Voltage Regulation
(1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
(1% of reading + 5V)
Short-Circuit Current
10mA max.
Output Impedance
2
RAMP TIME (Rise Time)
0.1s~999.9s
FALL Time
OFF~999.9s
WAIT TIME
OFF~999.9s
TIMER (Test Time)*
0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
GND
ON/OFF
Continuity Test
Output-Current
100mA dc
Ohmmeter Measurement Accuracy
1 ± 0.2, ON/OFF
Interface
Signal I/O
Standard
RS-232C
Standard
USB (device)
Standard
USB (host)
Standard (for parameter / LCD hardcopy)
Rear Output
Scanner
DISPLAY
4.3” color LCD
POWER SOURCE
AC 100V~240V ± 10%, 50Hz/60Hz
DIMENSION & WEIGHT
320(W) x 120(H) x 435(D) mm; Approx. 11kg