Data Sheet

GPT-9500 NPI Announcement_E 9
ISO-9001 CERTIFIED MANUFACTURER
GOOD WILL INSTRUMENT CO., LTD.
T
E
L
:
8
8
6
-
2
-
2
2
6
8
0
3
8
9
F
A
X
:
8
8
6
2
2
2
6
8
0
6
3
9
Specifications
AC Withstanding
Output-Voltage Range
0.050kV~5.000kV
Output-Voltage Resolution
1V
Output-Voltage Accuracy
(1% of setting + 5V) [no load]
Maximum Rated Load
150 VA (5kV/30mA)
Maximum Rated Current
30mA
0.001mA ~ 10mA (0.05kV V 0.5kV)
0.001mA ~ 30mA (0.5kV< V 5kV)
Output-Voltage Waveform
Sine wave
Voltage Regulation
(1% + 5V) [maximum rated load → no load]
Output-Voltage Frequency
50 Hz / 60 Hz selectable
Voltmeter Accuracy
(1% of reading + 5V)
Current Measurement Range
0.001mA ~ 30.00mA
Current Best Resolution
1μA (0.001mA ~ 9.999mA)
10μA (10.00mA ~ 30.00mA)
Current Measurement Accuracy
(1.5% of reading + 30μA)
Current Offset
80μA maximum
ARC Detect
Yes
RAMP TIME (Rise Time)
0.1s~999.9s
FALL Time
OFF~999.9s
WAIT Time
OFF~999.9s
TIMER (Test Time)
CONT
2
, 0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
GND
ON/OFF
DC Withstanding
Output-Voltage Range
0.050kV~6.000kV
Output-Voltage Resolution
1V
Output-Voltage Accuracy
(1% of setting + 5V) [no load]
Maximum Rated Load
50W (5kV/10mA)
Maximum Rated Current
10mA
0.001mA ~ 2mA (0.05kV V 0.5kV)
0.001mA ~ 10mA (0.5kV< V 6kV)
Voltmeter Accuracy
(1% of reading + 5V)
Voltage Regulation
(1% + 5V) [maximum rated load → no load]
Current Measurement Range
0.001mA ~ 10.00mA
Current Best Resolution
0.1μA (0.1μA ~ 999.9μA)
1μA (1μA ~ 9.999mA)
10μA (10.00mA)
Current Measurement Accuracy
(1% of reading + 1μA) when I Reading < 1mA
(1% of reading + 10μA) when I Reading 1mA
Current Offset
5μA maximum
ARC Detect
Yes
RAMP TIME (Rise Time)
0.1s~999.9s
FALL Time
OFF~999.9s
WAIT Time
OFF~999.9s
TIMER (Test Time)
CONT
2
, 0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
GND
ON/OFF
Insulation Resistance
Output Voltage
50V~1000V dc
Output-Voltage Resolution
1V
Output-Voltage Accuracy
± (1% of setting + 5V) [no load]
Resistance Measurement
0.1MΩ~ 10GΩ