User Manual Manual

AFG-2225 Series User Manual
194
*RST
System Command
Reset the function generator to its factory default
state.
Note the *RST command will not delete
instrument save states in memory.
*RST
SYSTem:VERSion?
System Query
Performs a system version query. Returns a string
with the instrument, firmware version, FPGA
revision
SYSTem:VERSion?
<string>
SYST:VERS?
AFG-2225 VX.XXX_XXXX
Returns the year and version for that year (1).
*OPC
System Command
This command sets the Operation Complete Bit
(bit 0) of the Standard Event Status Register after
the function generator has completed all pending
operations. For the AFG-2225, the *OPC command
is used to indicate when a sweep or burst has
completed.
Before the OPC bit is set, other commands may be
executed.
*OPC