COUNter:VALue?
Instrument
Command
Returns the current value from the frequency
counter.
COUNter:VALue?
COUNter:VALue?
+5.00E+02
Returns the frequency as 500Hz.
PHASE
The phase command remotely controls the phase and channel
synchronization.
SOURce[1|2]:PHASe
Instrument
Command
Sets the phase.
SOURce[1|2]:PHASe {<phase>|<MIN>|<MAX>}
-180~180
Sets the phase to the minimum
value.
Sets the phase to the maxium
value.
SOURce1:PHASe 25
Sets the phase of channel 1 to 25°.
SOURce[1|2]:PHASe? {MAX|MIN}
phase
Returns the current phase.
SOURce1:PHASe?
26
Returns the phase of channel 1 as 26°.