Triggers only on the AC component.
Triggers on AC+DC component.
LF
Puts a high-pass filter and rejects the
frequency below 50kHz.
HF
Puts a low-pass filter and rejects the
frequency above 50kHz.
Rejects noise signal.
Setup Holdoff and Auto level
The Holdoff function defines the waiting period
before the GBS-1000 starts triggering again after a
trigger point. The Auto level function
automatically adjusts the trigger level to the center
amplitude of the waveform.
1. Press the Trigger menu key
twice.
MENU
MENU
2. To set the Holdoff time,
press F1 (Holdoff) and use
the Variable knob. The
resolution depends on the
horizontal scale.
Holdoff
40.00ns
F 1
Range
40ns~2.5s
Pressing F2 (Set to
Minimum) sets the Holdoff
time to the minimum, 40ns.
Set To
Minimum
F 2
The holdoff function is automatically disabled
when the waveform update mode is in Roll or
Scan mode (page110).