Note
When using thermocouple wiring, please use
extension wires so that the cold junction points are
external to the scanner card. Connecting
thermocouple wiring directly to the scanner box is
not recommended due to the radiant heat from the
internal components.
Best if longer
than 20cm
Thermocouple
wire
Cold junction
points
Extension
wires
Hot junction
point
3. Bundle the wires at the front cover using the holes at
the bottom.
4. Close the top cover and tighten the screw from the
bottom.
Configuration
Record
Print out the configuration record list on page79, fill in