User Manual
REMOTE CONTROL
105
*ese <0~255> Sets ESER contents.
Example: *ese 65 sets ESER to 01000001
*esr? Returns and clears SESR (Standard Event Status Register).
Example: 198 means SESR=11000110
*idn? Returns company name, model No., and system version.
Example: GW, GDM8255A, 1.0
*opc? “1” is placed in the output queue when all the pending
operations are completed.
*opc Sets operation complete bit (bit0) in SERS (Standard Event
Status Register) when all pending operations are completed.
*psc? Returns power On clear status.
Parameter: 0 (cleared), 1 (not cleared)
*psc Clears power On status.
Parameter: 0 (clear), 1 (don’t clear)
*rst Recalls default panel setup (reset the device).
*sre? Returns SRER (Service Request Enable Register) contents.
Example: 3 means SRER=00000011
*sre <0~255> Sets SRER contents.
Example: *SRE 7 SRER=00000111
*stb? Returns SBR (Status Byte Register) contents.
Example: 81 means SBR=01010001
*trg Manually triggers the GDM-8200A series.
ROUTe command
rout:clos Close specified scanner channel.
Parameter: NR1, min, max
Example: rout:clos_102 (close channel102
rout:open:all Opens all scanner channels.
rout:mult:open Enable all channels in specified range.
Parameter: beginning channel, end channel
Example: rout:mult:open 105, 110 (105 to 110 enabled,
others disabled)
rout:mult:clos Disable all channels in specified range.
Parameter: beginning channel, end channel
Example: rout:mult:clos 105, 110 (105 to 110 disabled, others
enabled)
rout:mult:stat? Returns scanner box all channel status.
Parameter: 101 ON, 102 OFF, …. 201 ON, 202 OFF…