User Manual

BASIC MEASUREMENT
39
Set reference junction temperature
Background
W
hen a thermocouple is connected to the GDM-8200A
series, the temperature difference between the
thermocouple lead and the the GDM-8200A series input
terminal should be taken into account and be cancelled;
otherwise an erroneous temperature might be added.
Type Range Resolution
SIM
(simulated)
0 ~ +50°C 0.01°C
The terminal temperature is manually defined by the user.
Default value: 23.00
1. Open
reference
junction menu
Press the Shift key, the °C/°F
(Sensor) key, then the Down key.
The reference junction selection
menu appears on the display.
2. Edit reference
temperature
Use the Left/Right key to move the
cursor, and use the Up/Down key to
change the value.
Default: 23.00
Press the Enter key to confirm the
v
alue, or the Exit key to cancel. The
display goes back to the default
state.
(confirm)
(cancel)