User Manual

ADVANCED MEASUREMENT
51
5. Result High
If the 2
nd
display
shows High, the result
is above the High
limit.
Digital I/O: FAIL Out (Pin 6) and HIGH
Limit FAIL Out (Pin 7) are activated.
Low
If the 2
nd
display
shows Low, the result
is below the Low limit.
Digital I/O: FAIL Out (Pin 6) and LOW
Limit FAIL Out (Pin 8) are activated.
Pass
If the 2
nd
display
shows Pass, the result
is staying between the
High and the Low
limit.
Digital I/O: PASS Out (Pin 5) is activated.
Digital I/O
The Compare
measurement result
comes out from the
rear panel Digital I/O
terminal. For the
terminal details, see
page89.
Deactivate
Compare
measurement
To cancel the Compare
measurement, press the Shift
key followed by the HOLD
(Comp) key, or simply activate
another measurement.