Overview
The timer setting sets the “current” timer time
that is used to time stamp readings when
saving to USB. By default the timer time is the
elapsed time from when the unit was turned
on.
If the timer time ticks over 23:59:59, the timer
will revert back to 00:00:00 and the time stamp
will include a “day” count for each time this
occurs. Note, however, the “day” count cannot
be set in the timer settings.
Note
The GDM-834X uses volatile RAM and does not
have a CMOS backup battery to save the TIMER
settings when the power is turned off. When the
power is reset, the TIMER setting will be reset to
00:00:00.
Procedure
1. Press the MENU key.
2. Go to USBSTO on level 1.
3. Go to TIMER on level 2.
4. Set TIMER time between 00:00:00 and 23:59:59.
5. Press the Enter key to confirm.
6. Press the EXIT key to exit from the TIMER
menu.
Display
Timer setting
Timer menu
indicator