Manual

GDS-1000-U Programming Manual
46
Note
This command will not alter the position of the
waveform (RefA, W1~15) in memory, unless the
template is saved with the :TEMPlate:SAVe
:MAXimum command.
Before this command can be used, please set the
template mode to normal using
the :TEMPlate:MODe 0 command.
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:TEMPlate:POSition:MAX
<NR2>
:TEMP:POS:MAX?
:TEMP:POS:MAX
<NR2>
:TEMP:POS:MAX?
Parameter/
Return parameter
<NR2>
-12.00 ~ 12.00 Div. 0 represents the center
division.
Example
:TEMP:POS:MAX 2.00
Sets the template to
the 2
nd
grid division
above the center
division.
:TEMPlate:POSition:MIN
Set
Query
Description
Sets and queries the position of the MIN template
in grid divisions. 1 grid division = 25 on-screen
pixels.