COMMAND DETAILS
47
Note
This command will not alter the position of the
waveform (RefA, W1~15) in memory, unless the
template is saved with the :TEMPlate:SAVe
:MINimum command.
Before this command can be used, please set the
template mode to normal using
the :TEMPlate:MODe 0 command.
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:TEMPlate:POSition:MIN
<NR2>
:TEMP:POS:MIN?
:TEMP:POS:MIN
<NR2>
:TEMP:POS:MIN?
Parameter/
Return parameter
<NR2>
-12.00 ~ 12.00 Div. 0 represents the center
division.
Example
:TEMP:POS:MIN 2.00
Sets the template to
the 2
nd
grid division
above the center
division.
:TEMPlate:SAVe:MAXimum
Set
Description
Saves the maximum template.
Same as: Utility key More (F5) Go-NoGo
Menu(F1)Template Edit(F1)Save & Create(F4).
Note
Before this command can be used, please set the
template mode to normal using
the :TEMPlate:MODe 0 command.
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:TEMPlate:SAVe:MINimum
:TEMP:SAV:MIN