GDS-1000-U Programming Manual
48
:TEMPlate:SAVe:MINimum
Set
Description
Saves the minimum template.
Same as: Utility key More (F5) Go-NoGo
Menu(F1)Template Edit(F1)Save & Create(F4).
Note
Before this command can be used, please set the
template mode to normal using
the :TEMPlate:MODe 0 command.
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:TEMPlate:SAVe:MINimum
:TEMP:SAV:MIN
:TEMPlate:TOLerance
Set
Query
Description
Sets or queries the tolerance (as a percentage) of
the auto template.
Note
Before this command can be used, please set the
template mode to auto using the
:TEMPlate:MODe 1 command.
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:TEMPlate:TOLerance <NR2>
:TEMPlate:TOLerance?
:TEMP:TOL <NR2>
:TEMP:TOL?
Parameter/
Return parameter
<NR2>
4.0 ~ 40 (0.4% ~ 40%).
Example
:TEMP:TOL 10
Sets the tolerance to
10%.