Digital Storage Oscilloscope GDS-1000-U Series USER MANUAL GW INSTEK PART NO.
April 2013 edition This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will Corporation. The information in this manual was correct at the time of printing.
TABLE OF CONTENTS Table of Contents SAFETY INSTRUCTIONS .................................. 8 Safety Symbols ....................................................................... 8 Safety Guidelines .................................................................... 9 Power cord for the United Kingdom .................................... 12 GETTING STARTED ........................................ 13 Main Features ................................................................... 13 Panel Overview .....
GDS-1000-U Series User Manual Trigger key 3/5 ..................................................................... 37 Trigger key 4/5 ..................................................................... 37 Trigger key 5/5 ..................................................................... 38 Utility key 1/10 (Utility #1) .................................................. 38 Utility 2/10 (Utility #2) ........................................................ 39 Utility key 3/10 (Utility #3) ...............
TABLE OF CONTENTS Edit: Source .......................................................................... 69 Edit: Setup Parameters ........................................................ 69 Run Data logging ................................................................. 71 CONFIGURATION .......................................... 72 Acquisition ........................................................................ 72 Selecting the acquisition mode ............................................
GDS-1000-U Series User Manual Using the USB file utilities ................................................... 99 Quick Save (HardCopy) ................................................... 101 Save ................................................................................ 103 File type/source/destination.............................................. 103 Saving the panel settings ................................................... 104 Saving the waveform ................................................
TABLE OF CONTENTS INDEX ..........................................................
GDS-1000-U Series User Manual SAFETY INSTRUCTIONS This chapter contains important safety instructions that should be followed when operating and storing the oscilloscope. Read the following before any operation to ensure your safety and to keep the oscilloscope in the best condition. Safety Symbols These safety symbols may appear in this manual or on the oscilloscope. WARNING Warning: Identifies conditions or practices that could result in injury or loss of life.
SAFETY INSTRUCTIONS Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Safety Guidelines General Guideline CAUTION Make sure the BNC input voltage does not exceed 300V peak. Never connect a hazardous live voltage to the ground side of the BNC connectors. It might lead to fire and electric shock. Do not place heavy objects on the oscilloscope.
GDS-1000-U Series User Manual Power Supply WARNING Fuse WARNING Cleaning the oscilloscope Operation Environment AC Input voltage: 100 ~ 240V AC, 47 ~ 63Hz The power supply voltage should not fluctuate more than 10%. Connect the protective grounding conductor of the AC power cord to an earth ground. Fuse type: T1A/250V To ensure fire protection, replace the fuse only with the specified type and rating. Disconnect the power cord before replacing the fuse.
SAFETY INSTRUCTIONS (Pollution Degree) EN 61010-1:2001 specifies pollution degrees and their requirements as follows. The oscilloscope falls under degree 2. Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”. Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. Pollution degree 2: Normally only non-conductive pollution occurs.
GDS-1000-U Series User Manual Power cord for the United Kingdom When using the oscilloscope in the United Kingdom, make sure the power cord meets the following safety instructions.
GETTING STARTED GETTING STARTED The Getting started chapter introduces the oscilloscope’s main features, appearance, and set up procedure. Main Features Model name Frequency bandwidth Input channels GDS-1052-U DC –50MHz (–3dB) 2 GDS-1072-U DC – 70MHz (–3dB) 2 GDS-1102-U DC – 100MHz (–3dB) 2 Performance 250MSa /S real-time sampling rate 25GS/s equivalent-time sampling rate Up to 10ns peak detection 2mV~10V vertical scale 5.
GDS-1000-U Series User Manual Interface 14 USB 2.
GETTING STARTED Panel Overview Front Panel LCD Display Function keys V ARIABLE knob Vertical POSITION knob VOLTS/DIV knob Horizontal POSITION knob Menu keys VARIABLE Acquire Display Utility Cursor Measure Save / Recall VERTICAL Help Autoset Hardcopy Run / Stop HORIZONTAL TRIGGER Trigger LEVEL knob LEVEL CH 1 MATH CH 2 MENU VOLTS / DIV TIME / DIV MENU Horizontal MENU key SINGLE VOLTS / DIV FORCE CH 1 CH 2 X EXT TRIG Y Digital Storage Oscilloscope GDS - 1042-U GDS-1052-U
GDS-1000-U Series User Manual Utility key Utility Configures the Hardcopy function (page 101), shows the system status (page 94), selects the menu language (page 94), runs the self calibration (page 116), configures the probe compensation signal (page 117), and selects the USB host type (page 93). Help key Help Shows the Help contents on the display (page 44). Autoset key Autoset Automatically configures the horizontal, vertical, and trigger settings according to the input signal (page 46).
GETTING STARTED Horizontal menu key MENU Moves the waveform horizontally (page 78). Horizontal position knob TIME/DIV knob TIME/DIV VOLTS/DIV knob Input terminal CH 1 VOLTS/DIV CH1 Configures the vertical scale and coupling mode for each channel (page 82). Selects the vertical scale (page 82). Accepts input signals: 1MΩ±2% input impedance, BNC terminal. Accepts the DUT ground lead to achieve a common ground. Ground terminal MATH key Selects the horizontal scale (page 78).
GDS-1000-U Series User Manual Power switch 18 Powers the oscilloscope on or off.
GETTING STARTED Rear Panel Security lock slot Fuse socket Power cord socket CAL output LINE VOLTAGE RANGE FREQUENCY USB port AC 100 240V 50 60Hz T1A 250V FUSE RATING POWER MAX. 18W 40VA Power cord socket Power cord socket accepts the AC mains, 100 ~ 240V, 50/60Hz. Fuse socket The fuse socket holds the AC main fuse, T1A/250V. For the fuse replacement procedure, see page 122. USB slave port Accepts a type B (slave) male USB connector for remote control of the oscilloscope (page 93).
GDS-1000-U Series User Manual Display Waveform marker Waveform position Trigger status Acquisition Menu Vertical status Horizontal status Frequency Trigger condition Waveforms Channel 1: Yellow Channel 2: Blue Trigger status Trig’d A signal is being triggered Trig? Waiting for a trigger condition Auto Updating the input signal regardless of trigger conditions STOP Triggering is stopped For trigger setting details, see page 84.
GETTING STARTED Setting up the Oscilloscope Background This section describes how to set up the oscilloscope properly including adjusting the handle, connecting a signal, adjusting the scale, and compensating the probe. Before operating the oscilloscope in a new environment, run these steps to make sure the oscilloscope is functionally stable. Procedure 1. Pull both bases of the handle out slightly. 2. Turn to one of the three preset positions. 3. Connect the power cord. 4. Press the power switch.
GDS-1000-U Series User Manual 6. Connect the probe between the Channel1 input terminal and probe compensation signal output (2Vp-p, 1kHz square wave). 7. Set the probe attenuation voltage to x10. VARIABLE Acquire Display Utility Help Cursor Measure Save/Recall Hardcopy VERTICAL Autoset Run/Stop HORIZONTAL TRIGGER LEVEL CH 1 MATH CH 2 MENU VOLTS/DIV TIME/DIV MENU SINGLE VOLTS/DIV FORCE CH1 CH2 X EXT TRIG Y 1MW 300V 15pF CAT MAX. 300Vpk 1MW 300V 15pF CAT MAX.
GETTING STARTED 10. Turn the adjustment point on the probe to flatten the square waveform edge. Over Compensation Normal Under Compensation 11. Setting up the oscilloscope is complete. You may continue with the other operations.
GDS-1000-U Series User Manual QUICK REFERENCE This chapter lists the oscilloscope menu tree, operation shortcuts, built-in help coverage, and default factory settings. Use this chapter as a handy reference to access the oscilloscope functions.
QUICK REFERENCE Select acquisition mode Acquire Normal ~ Peak-Detect Select average number Normal Average 2/ 4/ 8/ 16/ 32/ 64/ 128/ 256 Average Turn Delay on/off Peak Detect Sample Rate 500MS/s CH1/CH2 key Turn channel on/off CH1 CH 1/2 Coupling Invert Off BW Limit Off Probe x1 / / On/Off On/Off On/Off (GDS-1102/1062-U) x1/x10/x100 Select coupling mode Coupling Invert waveform Invert Turn bandwidth limit on/off BW Limit Select probe attenuation ↔Probe 25
GDS-1000-U Series User Manual Cursor key 1/2 Turn cursor on/off Cursor Cursor Source CH1 CH1/ 2/ MATH Move X1 cursor X1→ VAR X1 123.4us 212.0mV X2 22.9us 0.000V Move X2 cursor X2→ VAR X1 X2 2 3 .6 u s 1 1 .9 H z 2 1 2 .0 mV Move both X1 and X2 cursor X1X2→ VAR X↔Y Switch to Y cursor X↔Y Cursor key 2/2 Turn cursor on/off Cursor Cursor Source CH1 CH1/ 2/ MATH Move Y1 cursor Y1 123.4mV Y1→ VAR Y2 12.9mV Move Y2 cursor Y1Y2 1 0 .
QUICK REFERENCE Display key Select waveform type Display Type Type Vectors Accumulate Vectors/ Dots On/ Off Off Waveform accumulate On/Off Accumulate Refresh accumulation Refresh Refresh Contrast Set display contrast Contrast →VAR Full / / Select display grid Autoset key Autoset Automatically find the signal and set the scale Autoset Hardcopy key Hardcopy → See Utility key (page 38) Help key Help Turn help mode on/off Help 27
GDS-1000-U Series User Manual Horizontal menu key Select main (default) display MENU Main Select window mode Main Window →TIME/DIV Window Zoom in window mode Window Zoom Window Zoom Select window roll mode Roll Roll XY Select XY mode XY Math key 1/2 (+/-) Math on/off MATH Operation CH1+CH2 CH1+CH2 CH1-CH2 FFT Math Select math operation type (+/– /FFT) Operation Position 0.
QUICK REFERENCE Math key 2/2 (FFT) Math on/off MATH Operation FFT Source CH1 Window Hanning Position 0.
GDS-1000-U Series User Manual Measure key Measure Measure Vpp 1:204mV 2: 300mV Vavg 1: 1.93mV 2: 28.0mV Frequency Voltage Vpp 1: 1.000kHz 2: 3.003kHz Voltage/Time Duty Cycle 1: 50.01% 2: 49.88% Rise Time Previous Menu 1. 76.20us 2. 70.
QUICK REFERENCE Save/Recall key 1/9 Save/Recall Save/Recall Default Setup Save Setup To Save Setup Save Waveform Recall Setup Recall Waveform Display Refs.
GDS-1000-U Series User Manual Save/Recall key 3/9 Select other menu Recall Waveform Recall Waveform Source Memory Destination Recall Waveform Memory/USB Select waveform source Ref A/B Source Select waveform destination Recall File Utilities →VAR (USB only) To File Utilities Destination →VAR Recall waveform Recall Go to USB file utilities File Utilities Save/Recall key 4/9 Select other menu Display Refs. Display Refs. Ref.A Off Ref.B Off Display Refs. On/ Off Turn ref.
QUICK REFERENCE Save/Recall key 5/9 Select other menu Save Setup Save Setup Save Setup Select destination Destination Memory Memory/USB File →VAR Save setup Save Utilities Destination (USB only) To File Utilities Save Go to USB file utilities File Utilities Save/Recall key 6/9 Select other menu Save Waveform Save Waveform Source Destination Memory Save Waveform CH1/2/Math Ref A/B Memory(W1~W15) USB Refs (Ref A, Ref B) Save File Utilities (USB only) To File Utilities Select source Source
GDS-1000-U Series User Manual Save/Recall key 7/9 Select other menu Save Image Save Image Ink Saver Off Save Image On/ Off Destination Turn on/off ink saver USB Ink Saver Save Save image File Utilities (USB only) To File Utilities Save Go to USB file utilities File Utilities 34
QUICK REFERENCE Save/Recall key 8/9 Select other menu Save All Save All Ink Saver Off Save All On/ Off Destination Turn on/off ink saver USB Ink Saver Save Select destination File Utilities (USB only) To File Utilities Destination →VAR Save all Save Go to USB file utilities File Utilities Save/Recall key 9/9 Select file/folder File Utilities Select Enter Character New Folder Back Space VAR →Select Create or rename folder/file New Folder/Rename Rename Delete Save Previous Menu Previ
GDS-1000-U Series User Manual Trigger key 1/5 Select Trigger type Trigger Type MENU Type Type Edge Source CH1 Slope / Coupling Mode Auto Trigger key 2/5 Select video trigger type Video Trigger Type Video Source CH1 Standard NTSC Polarity Line Type Select trigger source CH1/2 NTSC/SECAM/PAL / Field 1/ Field 2/ Line Source Select video standard Standard Select video polarity Polarity Select video field/line Line 36 →VAR
QUICK REFERENCE Trigger key 3/5 Select edge trigger type Edge Trigger Type Edge Source CH1 Edge CH1/2/Ext/Line Select trigger source Source Slope / Coupling Mode Auto To Slope/Coupling Auto/ Normal Go to slope/coupling menu (page 38) Slope/Coupling Select trigger mode Mode Trigger key 4/5 Select pulse trigger type Pulse Trigger Type Pulse Source CH1 When < 20.
GDS-1000-U Series User Manual Trigger key 5/5 Select trigger slope type Coupling/Slope Slope Coupling AC Rejection Off Noise Rej Off / AC/ DC LF/ HF/ Off On/ Off Previous Menu Slope Select trigger coupling mode Coupling Select frequency rejection Rejection Turn noise rejection on/off Noise Rej Go back to previous menu Previous Menu Utility key 1/10 (Utility #1) Go to hardcopy menu Utility Hardcopy Hardcopy Menu To Hardcopy menu Go to probe compensation menu ProbeComp Menu To Probe Comp menu Pr
QUICK REFERENCE Utility 2/10 (Utility #2) Go to the Go-NoGo menu Utility Go-NoGo Menu No Go When Data logging Menu More To Go-NoGo menu / To Data Logging menu To Utility #3 menu Go-NoGo Set the NoGo conditions to inside /outside limits No Go When Go to the Data Logging Menu Data Logging Go to the next Utility menu More 39
GDS-1000-U Series User Manual Utility key 3/10 (Utility #3) Enter self calibration Calibration Self CAL Menu To Self CAL menu Self CAL Go to the first Utility menu More To Utility #1 menu More Utility key 4/10 (Hardcopy -Save All) Select Hardcopy function Hardcopy Function Save All SaveImage/ SaveAll Function Ink Saver Off On/ Off Turn on/off Ink saver Ink Saver Previous Menu Go to previous menu Previous Menu Utility key 5/10 (Hardcopy -Save Image) Hardcopy- Save Image Select Hardcopy func
QUICK REFERENCE Utility key 6/10 (Probe compensation) Select probe compensation signal Probe compensation Wave Type / Wave Type Frequency 1K ( only) 1k ~ 100k Set frequency for square wave Duty Cycle ( only) 5% ~ 95% Frequency →VAR 50% Set duty cycle for square wave Default 1kHz Duty Cycle →VAR Previous Menu Go to previous menu Previous Menu Utility key 7/10 (Go-NoGo) Switch between templates Edit Template Max Source W 01 Tolerance 0.
GDS-1000-U Series User Manual Utility key 8/10 (Data Logging 1/2) Turn Data Logging On/Off Data logging Data logging Off Source CH1 Setup File Utilities Previous Menu On/Off Data logging CH1/CH2 Set the logging source To the Edit menu Source (USB only) To File Utilities Go to the Data Logging Edit menu To previous menu Setup Go to the File Utilities menu File Utilities Go back to previous menu Previous Menu Utility key 9/10 (Data Logging 2/2) Edit Save Waveform Interval 2 secs Duration 5 mins
QUICK REFERENCE Default Settings Here are the factory installed panel settings which appear when pressing the Save/Recall key→ Default Setup. Save/Recall Default Setup Acquisition Mode: Normal Channel Scale: 2V/Div Invert: Off Coupling: DC Probe attenuation voltage: x1 BW limit: Off Channel 1 & 2: On Cursor Source: CH1 Cursor: Off Display Type: Vectors Accumulate: Off Grid: Full Horizontal Scale: 2.5us/Div Mode: Main Timebase Math Type: + (Add) Position: 0.
GDS-1000-U Series User Manual Built-in Help The Help key shows the contents of the built-in help support. When you press a function key, its descriptions appear in the display. Applicable keys Acquire Display Cursor Measure Utility Save/Recall Hardcopy (Vertical) CH 1 Help Help Autoset Run/Stop (Horizontal) (Trigger) MATH CH 2 MENU MENU SINGLE FORCE Procedure 1. Press the Help key. The display changes to the Help mode. 2. Press a functional key to access its help contents.
MEASUREMENT MEASUREMENT The Measurement chapter describes how to properly observe a signal using the oscilloscope’s basic functions, and how to observe a signal in a detailed manner using some of the advanced functions such as: Automatic measurements, cursor measurements, and math operations. Basic Measurements This section describes the basic operations required in capturing and viewing an input signal. For more detailed operations, see the following chapters.
GDS-1000-U Series User Manual Channel 1 off Channel 1 on Channel indicator Channel icon De-activating a channel To de-activate the channel, press the Channel key twice (once if the channel menu is already selected). Using Autoset Background Procedure The Autoset function automatically configures the panel settings to the best viewing conditions, in the following way.
MEASUREMENT Before Autoset After Autoset Undo option Undoing the Autoset To undo the Autoset, press Undo (available for a few seconds). Adjusting the trigger level If the waveform is still unstable, try adjusting the trigger level up or down by using the Trigger Level knob. Limitation Autoset does not work in the following situation.
GDS-1000-U Series User Manual Trigger Run mode Trigger Stop mode Auto or Trig icon Waveform operation Stop icon Waveforms can be moved or scaled in both the Run and Stop mode. For details, see page 78 (Horizontal position/scale) and page 82 (Vertical position/scale). Changing the horizontal position and scale For more detailed configurations, see page 78. Setting the horizontal position The horizontal position knob moves the waveform left or right.
MEASUREMENT Selecting the horizontal scale To select the timebase (scale), turn the TIME/DIV knob; left (slow) or right (fast). Range Horizontal scale: 50us/div TIME/DIV 1ns/Div ~ 10s/Div, 1-2.5-5 increment Horizontal scale: 250us/div Changing the vertical position and scale For more detailed configuration, see page 82. Set vertical position To move the waveform up or down, turn the vertical position knob for each channel.
GDS-1000-U Series User Manual Using the probe compensation signal Background This section introduces how to use the probe compensation signal for general usage, in case the DUT signal is not available or to get a second signal for comparison. For probe compensation details, see page 117. Note: The frequency accuracy and duty factor are not guaranteed. Therefore the signal should not be used for reference purposes. Square waveform used for probe compensation. 1k ~ 100kHz, 5% ~ 95%.
MEASUREMENT 4. Press Wave type repeatedly to select the wave type. 5. (For only) To change the frequency, press Frequency and use the Variable knob. Range Probe compensation Frequency 1K VARIABLE 1kHz ~ 100kHz 6. (For only) To change the duty cycle, press Duty Cycle and use the Variable knob. Range Wave Type Duty Cycle 50% VARIABLE 5% ~ 95% For probe compensation details, see page 117.
GDS-1000-U Series User Manual Automatic Measurements The automatic measurement function measures input signal attributes and updates them in the display. Up to 5 automatic measurement items can be updated at any one time on the side menus. All automatic measurement types can be displayed on screen if necessary.
MEASUREMENT Time measurement items Vhi Global high voltage. Vlo Global low voltage. Vavg Averaged voltage of the first cycle. Vrms RMS (root mean square) voltage. ROVShoot Rise overshoot voltage. FOVShoot Fall overshoot voltage. RPREShoot Rise preshoot voltage. FPREShoot Fall preshoot voltage. Frequency of the waveform. Freq T Period Waveform cycle time (=1/Freq). Risetime Rising time of the pulse (~90%). Falltime Falling time of the pulse (~10%). +Width Positive pulse width.
GDS-1000-U Series User Manual Automatically measuring the input signals Viewing the measurement result 1. Press the Measure key. Measure 2. The measurement results appear on the menu bar, constantly updated. 5 measurement slots (F1 to F5) can be customized. Selecting a measurement item 3. Press F3 repeatedly to select the measurement type: Voltage or Time. 4. Use the Variable knob to select the measurement item. 5.
MEASUREMENT Cursor Measurements Cursor line, horizontal or vertical, shows the precise position of the input waveforms or the math operation results. The horizontal cursor can track time, voltage and frequency, whilst the vertical cursor can track voltage. Using the horizontal cursors Procedure 1. Press the Cursor key. The cursors appear in the display. Cursor 2. Press X↔Y to select the horizontal (X1&X2) cursor. X↔Y 3. Press Source repeatedly to select the source channel.
GDS-1000-U Series User Manual To move the right cursor, press X2 and then use the Variable knob. Remove cursors X2 22.9us 0.000V To move both cursors at once, press X1X2 and then use the Variable knob. X1X2 23.6us 11.9Hz 212.0mV Press Cursor to remove the onscreen cursors. Cursor Using the vertical cursors Procedure 1. Press the Cursor key. 2. Press X↔Y to select the vertical (Y1&Y2) cursor. 3. Press Source repeatedly to select the source channel. Range Cursor X↔Y Source CH1 CH1, 2, MATH 4.
MEASUREMENT To move both cursors at once, press Y1Y2 and then use the Variable knob. Remove cursors Press Cursor to remove the onscreen cursors. Y1Y2 10.
GDS-1000-U Series User Manual Math Operations The Math operations can add, subtract, or perform FFT on the input waveforms. The resulted waveform can be measured using the cursors, and saved or recalled just like normal input signals. Overview Addition (+) Adds the amplitude of CH1 & CH2 signals. Subtraction (–) Extracts the amplitude difference between CH1 & CH2. FFT Performs a FFT calculation on a signal. Four types of FFT windows are available: Hanning, Flattop, Rectangular, and Blackman.
MEASUREMENT Adding, subtracting or multiplying signals Procedure 1. Activate both CH1 and CH2. CH 1 2. Press the Math key. MATH 3. Press Operation repeatedly to select addition (+) or subtraction (–). 4. The math measurement result appears in the display. 5. To move the math result vertically, use the Variable knob. The position will be displayed in Position. 6. To clear the math result from the display, press the Math key again. CH 2 Operation CH1+CH2 Unit/Div 2V Position 0.
GDS-1000-U Series User Manual 4. Press Window repeatedly to select the FFT window type. Window Hanning 5. The FFT result appears. The horizontal scale changes from time to frequency, and the vertical scale from voltage to dB. 6. To move the FFT waveform vertically, press Position and use the Variable knob. Range Unit/Div 1dB 1, 2, 5, 10, 20 dB/Div 8. To clear the FFT result from the display, press the Math key again. 60 VARIABLE –12.00 Div ~ +12.00 Div 7.
MEASUREMENT Go No-Go Testing Overview Background Go-NoGo testing checks if a waveform conforms to a user-specified maximum and minimum boundary (template). The testing can be set to stop or continue each time the template has or has not been violated by the input waveform.
GDS-1000-U Series User Manual Edit: NoGo When Procedure 1. Press the Utility key. Utility 2. Press the More key. More 3. Press No Go When repeatedly to select the NoGo conditions. No Go When NoGo when the waveform is inside the boundary (template) NoGo when the waveform is outside of the boundary (template) Edit: Source Procedure 1. Press the Utility key. Utility 2. Press the More key. More 62 3. Press the Go-NoGo Menu key. Go-NoGo Menu 4.
MEASUREMENT Edit: NoGo Violation Conditions Procedure 1. Press the Utility key. Utility 2. Press the More key. More 3. Press the Go-NoGo Menu key. Go-NoGo Menu 4. Press Violating repeatedly to select the NoGo conditions. Violating Stop Stop Stops the test when the NoGo conditions have been met. Continue The tests continue even when the NoGo conditions have been met. Edit: Template (boundary) Background The NoGo template sets the upper and lower amplitude boundary.
GDS-1000-U Series User Manual customizable. Disadvantage: The waveforms (templates) have to be stored internally prior to this selection. Auto Creates the upper and lower boundary (template) from the source signal, not from an internally stored waveform. Advantage: No need to store the waveforms prior to this selection. Disadvantage: The template shape is proportional to the source signal. The distance (allowance) between the source signal and the upper and lower template is the same. Max/Mix 1.
MEASUREMENT 7. Press Source and use the Variable knob to select the waveform template. Source W 01 VARIABLE Max Waveform A: Ref A, W01~W15 Min Waveform B: Ref B, W01~W15 8. Press Position and use the Variable knob to set the waveform amplitude. Source W 01 VARIABLE 9. Repeat steps 5-7 for the other template setting (Max or Min). 10. When both Max and Min templates have been configured, press Save & Create to save the templates.
GDS-1000-U Series User Manual Auto 1. The template is based on the source signal. Ensure the source signal appears on the display. 2. Press the Utility key. Utility 3. Press the More key. More 4. Press the Go-NoGo Menu key. 5. Press the Template Edit key. 6. Press Template repeatedly to select the Auto template. 7. Press Source and use the Variable knob to select the template source. Go-NoGo Menu Template Edit Template Auto Source CH1 VARIABLE Source CH1, CH2 8.
MEASUREMENT Div 0.04 Div ~ 4.0 Div 9. When the Auto template has been configured, press Save & Create to save the template. Save & Create Auto template Tolerance Source waveform Run Go-NoGo Tests Procedure 1. Press the Utility key. Utility 2. Press the More key. More 3. Press the Go-NoGo Menu key. Go-NoGo Menu Ensure the source signal and boundary templates appear on the screen.
GDS-1000-U Series User Manual 4. Press Go-NoGo. The test Go-NoGo starts and stops according to On the conditions set on page 62, 63. To stop the test that has already started, press Go-NoGo again. 5. The test results appear in the Ratio soft-key. The numerator denotes the total number of failed tests. The denominator denotes the total number of tests. Ratio: 2 BMP 9 BMP Numerator Number of “failed“ tests. Denominator Total number of tests.
MEASUREMENT Edit: Source Procedure 1. Press the Utility key. Utility 2. Press the More key. More 3. Press the Data logging Menu key. 4. Press Source repeatedly to select the source channel (CH1 or CH2). Data logging Menu Source CH1 Edit: Setup Parameters Background The logging function must set the type of data that will be logged (waveform/image), the capture interval time and the duration of the data logging. Procedure 1. Press the Utility key. Utility 2. Press the More key. More 3.
GDS-1000-U Series User Manual 5. Press Save repeatedly to log data or screen images. 6. Press Interval and use the Variable knob to select the interval time. Save Waveform Interval 2 mins VARIABLE Interval time 2 secs~ 2min (duration = 5 min) 2 secs~ 5 min (duration 5~ 30 min) 2 secs~ 30 min (duration 30+ min) 7. Press Duration and use the Variable knob to set the duration time. Duration 5 mins VARIABLE Duration 5 mins ~ 100 hours 8. Press Previous menu to return to the Data logging menu.
MEASUREMENT Run Data logging Background Ensure the data source (page 69) and data logging setup has been set (page 69). Procedure 1. Insert a USB flash drive into the USB front panel port. 2. Press the Utility key. Utility 3. Press the More key. More 4. Press the Data logging Menu key. Data logging Menu 5. Press Data logging to turn Data logging data logging On. On Data/image files start logging to the USB flash drive automatically. To stop the Data logging, press the Data logging key again.
GDS-1000-U Series User Manual CONFIGURATION The Configuration chapter describes how to configure panel settings to make measurements and observations suited to the application needs. Acquisition The acquisition process samples the analog input signals and converts them into digital format for internal processing. You may select the normal, average, or peak detect acquisition mode. Selecting the acquisition mode Procedure 1. Press the Acquire key. 2.
CONFIGURATION Average Multiple data is averaged to form a waveform. This mode is useful for drawing a noise-free waveform. To select the number, press Average repeatedly. Average number: 2, 4, 8, 16, 32, 64, 128, 256 Peak detect To activate the Peak detect mode, press Peak-Detect. Only the minimum and maximum value pairs for each acquisition interval (bucket) are used. This mode is useful for catching abnormal glitches in a signal. Peak detect effect 1.
GDS-1000-U Series User Manual 7. Press Normal. Normal 8. Press Peak-Detect and see that a spike noise is captured. Example The peak detect mode reveals the occasional glitch.
CONFIGURATION Real time vs Equivalent time sampling mode Backgrounds The oscilloscope automatically switches between two sampling modes, Real-time and Equivalenttime, according to the number of active channels and sampling rate. Real-time sampling One sampled data is used to reconstruct a single waveform. Short-time events might get lost if the sampling rate gets too high. This mode is used when the sampling rate is relatively low (250MSa/s or lower).
GDS-1000-U Series User Manual Display The Display section describes how to configure the display settings: drawing type, waveform accumulation, contrast adjustment, and grid settings. Selecting vector or dot drawing Procedure 1. Press the Display key. 2. Press Type repeatedly to select the waveform drawing. Types Display Type Vectors Dots Only the sampled dots are displayed. Vectors The sampled dots are connected by lines.
CONFIGURATION Example Accumulation off Accumulation on Adjusting the display contrast Display Procedure 1. Press the Display key. 2. Press Contrast. Turn the Variable knob left to lower the contrast (dark display) or right to raise the contrast (bright display). Contrast VARIABLE Selecting the display grid Procedure 1. Press the Display key. Display 2. Press the grid icon repeatedly to select the grid. Full Parameters Shows the full grid. Shows the outer frame and X/Y axis.
GDS-1000-U Series User Manual Horizontal View The Horizontal view section describes how to configure the horizontal scale, position, waveform update mode, window zoom, and X-Y mode. Moving the waveform position horizontally Procedure The horizontal position knob moves the waveform left or right. The position indicator at the top of the display shows the center and current position.
CONFIGURATION Selecting the waveform update mode Background The display update mode is switched automatically or manually according to the horizontal scale. Main mode Updates the whole displayed waveform at once. The main mode is automatically selected when the horizontal scale (timebase) is fast. Roll mode Horizontal scale ≤100ms/div Trigger All modes available Updates and moves the waveform gradually from the right side of the display to the left.
GDS-1000-U Series User Manual Zooming the waveform horizontally Procedure/ range 1. Press the Horizontal Menu key. 2. Press Window. MENU Window 3. Use the horizontal position knob to move the zoom range sideways, and TIME/DIV knob to change the zoom range width. TIME/DIV The width of the bar in the middle of the display is the actual zoomed area. Zoom range 1ns ~ 25s 4. Press Window Zoom. The specified range gets zoomed.
CONFIGURATION Viewing waveforms in the X-Y mode Background The X-Y mode compares the voltage of Channel 1 and Channel 2 waveforms in a single display. This mode is useful for observing the phase relationship between the two waveforms. Procedure 1. Connect the signals to Channel 1 (X-axis) and Channel 2 (Y-axis). 2. Make sure both Channel 1 and 2 are activated. 3. Press the Horizontal key. CH 1 MENU 4. Press XY. The display shows two waveforms in X-Y format; Channel 1 as X-axis, Channel 2 as Y-axis.
GDS-1000-U Series User Manual Vertical View (Channel) The Vertical view section describes how to set the vertical scale, position, bandwidth limitation, coupling mode, and attenuation. Moving the waveform position vertically Procedure To move the waveform up or down, turn the vertical position knob for each channel. Selecting the vertical scale Procedure To change the vertical scale, turn the VOLTS/DIV knob; left (down) or right (up).
CONFIGURATION AC coupling mode. Only the AC portion of the signal appears on the display. This mode is useful for observing AC waveforms mixed with DC components. Inverting the waveform vertically Procedure 1. Press the Channel key. CH 1 2. Press Invert. The waveform becomes inverted (upside down) and the Channel indicator in the display shows a down arrow.
GDS-1000-U Series User Manual 2. Press BW Limit to turn on or BW Limit off the limitation. When Off turned on, the BW indicator appears next to the Channel indicator in the display. Example BW Limit Off BW Limit On Selecting the probe attenuation level Background A signal probe has an attenuation switch to lower the original DUT signal level to the oscilloscope input range, if necessary.
CONFIGURATION Trigger The Trigger function configures the conditions by which the oscilloscope captures the incoming signals. Trigger type Edge Triggers when the signal crosses an amplitude threshold in either a positive or negative slope. Video Extracts a sync pulse from a video format signal and triggers on a specific line or field. Pulse Triggers when the pulse width of the signal matches the trigger settings.
GDS-1000-U Series User Manual The Auto trigger status appears in the upper right corner of the display. Single The oscilloscope acquires the input signals once when a trigger event occurs, then stops acquiring. Pressing the Single key again will repeat the process. SINGLE The Single trigger status appears in the upper right corner of the display. (Searching) Normal (Triggered) The oscilloscope acquires and updates the input signals only when a trigger event occurs.
CONFIGURATION > Longer than = Equal to < Shorter than ≠ Not equal to Triggers on the rising edge. Trigger slope Triggers on the falling edge. Trigger coupling Frequency rejection AC Triggers only on AC component. DC Triggers on AC+DC component. LF Puts a high-pass filter and rejects the frequency below 50kHz. HF Puts a low-pass filter and rejects the frequency above 50kHz. Noise rejection Rejects noise signals.
GDS-1000-U Series User Manual Range Auto, Normal 5. Press Slope/coupling to enter into the trigger slope and coupling selection menu. 6. Press Slope repeatedly to select the trigger slope, rising or falling edge. Range Noise Rej Off On, Off 10. Press Previous menu to go back to the previous menu. 88 Rejection Off LF, HF, Off 9. Press Noise Rej to turn the noise rejection on or off. Range Coupling AC DC, AC 8. Press Rejection to select the frequency rejection mode.
CONFIGURATION Configuring the video trigger Procedure 1. Press the Trigger menu key. MENU 2. Press Type repeatedly to select video trigger. The video trigger indicator appears at the bottom of the display. Type Video 3. Press Source repeatedly to select the trigger source channel. Range Source CH1 Channel 1, 2 4. Press Standard repeatedly to select the video standard. Range Standard NTSC NTSC, PAL, SECAM 5. Press Polarity repeatedly to select the video signal polarity.
GDS-1000-U Series User Manual Configuring the pulse width trigger Procedure 1. Press the Trigger menu key. 2. Press Type repeatedly to select pulse width trigger. The pulse width trigger indicator appears at the bottom of the display. 3. Press Source repeatedly to select the trigger source. Range Type Pulse Source CH1 Channel 1, 2, Ext 4. Press Mode repeatedly to select the trigger mode, Auto or Normal. To select the Single trigger mode, press the Single key.
CONFIGURATION 7. Press Slope repeatedly to select the trigger slope, which also appears at the bottom of the display. Range Rising edge, falling edge 8. Press Coupling repeatedly to select the trigger coupling. Range Rejection Off LF, HF, Off 10. Press Noise Rej to turn the noise rejection on or off. Range Coupling AC DC, AC 9. Press Rejection to select the frequency rejection mode. Range Slope Noise Rej Off On, Off 11. Press Previous menu to go back to the previous menu.
GDS-1000-U Series User Manual To acquire the signal regardless of trigger conditions To acquire the input signal regardless of the trigger condition, press the Force key. The oscilloscope captures the signals once. FORCE In the Single trigger mode Press the Single key to start waiting for the trigger condition. To break out of the Single mode, press the Run/Stop key. The trigger mode changes to the Normal mode.
CONFIGURATION Rear Panel USB Port Interface The Remote control interface section describes how to set up the USB interface for PC connection. The details of remote control commands are described in the GDS-1000-U Programming Manual. USB connection Procedure PC end Type A, host GDS-1000-U end Type B, slave Speed 1.1/2.0 (full speed) 1. Connect the USB cable to the USB slave port on the GDS-1000-U. 2. When the PC asks for the USB driver, select dso_cdc_1000.
GDS-1000-U Series User Manual System Settings The system settings show the oscilloscope’s system information and allow changing the language. Viewing the system information Procedure 1. Press the Utility key. Utility 2. Press System Info. The upper half of the display shows the following information. System Info. Manufacturer Model Serial number Firmware version Web address 3. Press any other key to go back to the waveform display mode.
CONFIGURATION Procedure 1. Press the Utility key. Utility 2. Press Language repeatedly to Language select the language.
GDS-1000-U Series User Manual SAVE/RECALL The save function allows saving display images, waveform data, and panel settings into the oscilloscope’s internal memory or to the front panel USB port. The recall function allows recalling the default factory settings, waveform data, and panel settings from the oscilloscope’s internal memory or from USB. File Structures Three types of file are available: display image, waveform file, and panel settings. Display image file format Format xxxx.
SAVE/RECALL Waveform data format External USB Flash drive A USB flash drive (FAT or FAT32 format) can hold practically an unlimited number of waveforms. Ref A, B Two reference waveforms are used as a buffer to recall a waveform in the display. You have to save a waveform into internal memory or to USB, then copy the waveform into the reference waveform slot (A or B), and then recall the reference waveform into the display. One division includes 25 points of horizontal and vertical data.
GDS-1000-U Series User Manual Setup file format Format xxxx.set (proprietary format) A setup file saves or recalls the following settings.
SAVE/RECALL Using the USB file utilities Background When a USB flash drive is inserted into the oscilloscope, file utilities (file deletion, folder creation and file/folder renaming) are available from the front panel. Procedure 1. Insert a USB flash drive into the front panel USB port. 2. Press the Save/Recall key. Select any save or recall function. For example USB Destination in the Save image function. Save/Recall (Example) Save Image Destination USB 3. Press File Utilities.
GDS-1000-U Series User Manual Creating a new 1. Move the cursor to the file folder / renaming or folder location and press a file or folder New Folder or Rename. The file/folder name and the character map will appear on the display. 2. Use the Variable knob to move the pointer to the characters. Press Enter Character to add a character or Back Space to delete a character. 3. When editing is complete, press Save. The new/renamed file or folder will be saved. Deleting a folder 1.
SAVE/RECALL Quick Save (HardCopy) Background The Hardcopy key works as a shortcut to save display images, waveform data, and panel settings onto a USB flash drive card. Hardcopy The Hardcopy key can be configured into two types of operations: save image and save all (image, waveform, setup). Using the Save/Recall key can also save files with more options. For details, see page 103. Functionalities Procedure Save/Recall Save image (*.bmp) Saves the current display image into a USB flash drive.
GDS-1000-U Series User Manual 5. To invert the color in the display image, press Ink Saver. This turns Ink Saver on or off. 6. Press the Hardcopy key. The file or folder will be saved to the root directory of the USB flash drive.
SAVE/RECALL Save This section describes how to save data using the Save/Recall menu. File type/source/destination Item Source Panel setup (xxxx.set) Waveform data (xxxx.csv) Panel settings Destination Internal memory: S1 ~ S15 External memory: USB Channel 1, 2 Math operation result Internal memory: W1 ~ W15 Reference waveform A, B Reference waveform A, B External memory: USB Display image (xxxx.
GDS-1000-U Series User Manual Saving the panel settings Procedure 1. (For saving to USB flash drive) Insert the USB flash drive into the front panel USB port. 2. Press the Save/Recall key twice to access the Save menu. 3. Press Save Setup. Save/Recall Save/Recall Save Setup 4. Press Destination repeatedly Destination to select the saved location. Memory Use the Variable knob to VARIABLE change the internal memory location (S1 ~ S15).
SAVE/RECALL File utilities To edit the USB drive contents (create/ delete/ rename files and folders), press File Utilities. For details, see page 99. File Utilities Saving the waveform Procedure 1. (For saving to USB flash drive) Insert the USB flash drive into the front panel USB port. 2. Press the Save/Recall key twice to access the Save menu. 3. Press Save Waveform. 4. Press Source. Use the Variable knob to select the source signal.
GDS-1000-U Series User Manual Memory Internal memory, W1 ~ W15 USB Save to the USB flash drive with a 4k waveform memory length. Ref Internal reference waveform, A/B 6. Press Save to confirm saving. When completed, a message appears at the bottom of the display. Note File utilities Save The file will not be saved if the power is turned off or the USB flash drive is removed from the USB port. To edit the USB drive contents (create/ delete/ rename files and folders), press File Utilities.
SAVE/RECALL 4. Press Ink Saver repeatedly to Ink Saver invert the background color Off (on) or not (off). Note: Destination is set as USB. This cannot be changed. 5. Press Save to confirm saving. When completed, a message appears at the bottom of the display. Note File utilities Destination USB Save The file will not be saved if the power is turned off or the USB flash drive is removed before completion. To edit the USB drive contents (create/ delete/ rename files and folders), press File Utilities.
GDS-1000-U Series User Manual Saving all (panel settings, display image, waveform) Procedure 1. (For saving to USB flash drive) Insert the USB flash drive into the front panel USB port. 2. Press the Save/Recall key twice to access the Save menu. 3. Press Save All. The following information will be saved. Save/Recall Save/Recall Save All Setup file (Axxxx.set) Two types of setups are saved: the current panel setting and the last internally saved settings (one of S1 ~ S15). Display image (Axxxx.
SAVE/RECALL 6. Press Save to confirm saving. When completed, a message appears at the bottom of the display. Note Save The file will not be saved if the power is turned off or the USB flash drive is removed from the USB port. It takes approximately 1 min to save a 2M waveform to the USB drive in fast mode. Detailed mode may take over 10 times longer depending on the speed of the USB flash drive. 7.
GDS-1000-U Series User Manual Recall File type/source/destination Item Source Default panel setup Factory installed setting Current front panel Reference waveform Internal memory: A, B Current front panel Panel setup (DSxxxx.set) Internal memory: S1 ~ S15 Current front panel External memory: USB flash drive Internal memory: W1 ~ W15 Reference waveform A, B External memory: USB flash drive Waveform data (DSxxxx.
SAVE/RECALL Recalling the default panel settings Procedure 1. Press the Save/Recall key. 2. Press Default Setup. The factory installed setting will be recalled. Setting contents Acquisition Channel Cursor Display Horizontal Math Measure Trigger Utility Save/Recall Default Setup The following is the default panel setting contents. Mode: Normal Coupling: DC Invert: Off Probe attenuation: x1 BW limit: Off Source: CH1 Horizontal: None Vertical: None Type: Vectors Accumulate: Off Graticule: Scale: 2.
GDS-1000-U Series User Manual Recalling a reference waveform to the display Procedure The reference waveform must be stored in advance. See page 105 for details. 1. Press the Save/Recall key. 2. Press Display Refs. The reference waveform display menu appears. Save/Recall Display Refs. 3. Select the reference Ref.A Off waveform, Ref A or Ref B, and press it. The waveform appears on the display and the period and amplitude of the waveform appears in the menu. Ref.A On 1V 2.5ms 4.
SAVE/RECALL 3. Press Recall Setup. 4. Press Source repeatedly to select the file source, internal or external memory. Use the Variable knob to change the memory. Source Memory VARIABLE Memory Internal memory, S1 ~ S15 USB USB flash drive, DSXXXX.SET. The setup file(s) must be placed in the root directory to be recognized. 5. Press Recall to confirm recalling. When completed, a message appears at the bottom of the display.
GDS-1000-U Series User Manual 3. Press Recall Waveform. The display shows the available source and destination options. 4. Press Source repeatedly to select the file source, internal memory or USB. Use the Variable knob to change the memory location (W1 ~ W15)/DSXXXX.CSV. Source Memory VARIABLE Memory Internal memory, W1 ~ W15 USB USB flash drive, DSXXXX.CSV. The waveform file(s) must be placed in the root directory to be loaded. 5. Press Destination.
SAVE/RECALL File utilities To edit the USB drive contents (create/ delete/ rename files and folders), press File Utilities. For details, see page 99.
GDS-1000-U Series User Manual MAINTENANCE Two types of maintenance operations are available: calibrating the vertical resolution, and compensating the probe. Run these operations when using the oscilloscope in a new environment. Vertical Resolution Calibration Procedure 1. Press the Utility key. Utility 2. Press the More key twice. More x2 3. Press Self Cal Menu. 4. Press Vertical. The message “Set CAL to CH1, then press F5” appears at the bottom of the display. Self CAL Menu Vertical 5.
MAINTENANCE 7. The Channel1 calibration will complete in less than 5 minutes. 8. When finished, connect the calibration signal to the Channel 2 input and repeat the procedure. Ch1 calibration 1/3 Done!! 9. When the calibration is complete the display will go back to the previous state. Probe Compensation Procedure 1. Connect the probe between the Channel1 input and the probe compensation output (2Vp-p, 1kHz square wave) on the front panel. Set the probe voltage attenuation to x10.
GDS-1000-U Series User Manual 4. Press Wavetype repeatedly to Wave Type select the standard square wave. 5. Press the Autoset key. The compensation signal will appear in the display. Autoset 6. Press the Display key, then Type to select the vector waveform. Display Type Vectors 7. Turn the adjustment point on the probe until the signal edge becomes sharp.
FAQ FAQ • The input signal does not appear in the display. • I want to remove some contents from the display. • The waveform does not update (frozen). • The probe waveform is distorted. • Autoset does not catch the signal well. • Autoset function cannot catch signals under 30mV or 30Hz. Please use the manual operation. See page46 for details. I want to clean up the cluttered panel settings. • The accuracy does not match the specifications. The input signal does not appear in the display.
GDS-1000-U Series User Manual The waveform does not update (frozen). Press the Run/Stop key to unfreeze the waveform. See page 47 for details. For trigger setting details, see page 84. If this does not help, press the CH key. If the signal still does not appear, press the Autoset key. The probe waveform is distorted. You might need to compensate the probe. For details, see page 117.
FAQ The accuracy does not match the specifications. Make sure the device is powered on for at least 30 minutes, within +20°C~+30°C. This is necessary to stabilize the unit to match the specification. For more information, contact your local dealer or GWInstek at www.gwinstek.com / marketing@goodwill.com.tw.
GDS-1000-U Series User Manual APPENDIX Fuse Replacement Procedure 1. Remove the power cord and remove the fuse socket using a minus driver. Replace the fuse in the holder.
APPENDIX GDS-1000-U Series Specifications The specifications apply when the oscilloscope is powered on for at least 30 minutes under +20°C~+30°C. Model-specific specifications GDS-1052-U GDS-1072-U GDS-1102-U Bandwidth (–3dB) DC coupling: DC ~ 50MHz AC coupling: 10Hz ~ 50MHz Bandwidth Limit 20MHz (−3dB) Trigger Sensitivity 0.5div or 5mV (DC ~ 25MHz) 1.5div or 15mV (25MHz~50MHz) External Trigger ~ 50mV (DC~25MHz) Sensitivity ~ 100mV (25MHz~50MHz) Rise Time < 7ns approx.
GDS-1000-U Series User Manual Common specifications Vertical Trigger External trigger Horizontal X-Y Mode Signal Acquisition 124 Sensitivity Accuracy Bandwidth Rise Time Input Coupling Input Impedance Polarity Maximum Input Math Operation Offset Range 2mV/div~10V/Div (1-2-5 increments) ± (3% x |Readout|+0.1div + 1mV) See model-specific specifications See model-specific specifications AC, DC, Ground 1MΩ±2%, ~15pF Normal, Invert 300V (DC+AC peak), CAT II +, –, FFT 2mV/div~50mV/div: ±0.
APPENDIX Cursors and Measurement Control Panel Function Display Interface Probe Compensation Signal Power Source Operation Environment Storage Environment Dimensions Weight Voltage Vpp, Vamp, Vavg, Vrms, Vhi, Vlo, Vmax, Vmin, Rise Preshoot/ Overshoot, Fall Preshoot/ Overshoot Time Freq, Period, Rise Time, Fall Time, + Width, – Width, Duty Cycle Cursors Voltage difference (ΔV) and Time difference (ΔT) between cursors Auto Counter Resolution: 6 digits, Accuracy: ±2% Signal source: All available trigg
GDS-1000-U Series User Manual Probe Specifications GDS-1052-U & GDS-1072-U Probe Applicable model & GDS-1052-U, GDS-1072-U probe GTP-070A-4 Position x 10 Attenuation Ratio 10:1 Bandwidth DC ~ 70MHz Input Resistance 10MΩ when used with 1MΩ input Input Capacitance 28~32pF approx. Maximum Input ≤600Vpk, Derating with frequency Voltage Position x 1 Attenuation Ratio 1:1 Bandwidth DC ~ 6MHz Input Resistance 1MΩ when used with 1MΩ input Input Capacitance 120~220pF approx.
APPENDIX GDS-1102-U Probe Applicable model & GDS-1102-U probe GTP-100A-4 Position x 10 Attenuation Ratio 10:1 Bandwidth DC ~ 100MHz Input Resistance 10MΩ when used with 1MΩ input Input Capacitance 14.5~17.5pF approx. Maximum Input ≤600Vpk, Derating with frequency Voltage Position x 1 Attenuation Ratio 1:1 Bandwidth DC ~ 6MHz Input Resistance 1MΩ when used with 1MΩ input Input Capacitance 85~115pF approx. Maximum Input ≤200Vpk, Derating with frequency Voltage Operating Cond.
GDS-1000-U Series User Manual Dimensions 341.5 159.0 115.1 VARIABLE Acquire Display Utility Help Cursor Measure Save/ Recall Hardcopy VERTICAL Autoset Run/ Stop HORIZONTAL TRIGGER CH 2 MENU VOLTS/DIV TIME/DIV MENU SINGLE VOLTS/DIV 162.3 MATH 142.0 LEVEL CH1 FORCE CH1 CH2 X GDS-1042-U GDS-1052-U 40MHz 1GS/as 1MW 300V 15pF CAT MAX. 300Vpk 310.0 128 EXT TRIG Y Digital Storage Oscilloscope 1 MW 300V 15pF CAT MAX. 300Vpk 149.
APPENDIX EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan GOOD WILL INSTRUMENT (SUZHOU) CO., LTD. No.
GDS-1000-U Series User Manual INDEX AC coupling ............................... 83 Accumulating waveform.......... 76 Acquisition ................................. 72 menu tree ....................................... 25 specification ................................. 124 Addition ..................................... 59 Amplitude measure .................. 52 auto set faq.......................................... 119, 120 Auto set ....................................... 46 exception .....................
INDEX Falling time measure ................. 53 Faq ............................................. 120 FFT ............................................... 59 menu tree ....................................... 29 overview ........................................ 58 File format................................... 96 Firmware version ....................... 94 Flattop window .......................... 58 Force trigger ............................... 92 frequency cursor tracking ..............................
GDS-1000-U Series User Manual setup ............................................. 112 waveform ..................................... 113 Rectangular window ................. 58 Reference waveform menu tree ....................................... 32 recall ............................................. 112 Remote control interface........... 93 Rising time measure .................. 53 roll mode .................................... 79 Root mean square measure ...... 53 Run/stop .............................
INDEX 133