User Manual User Manual
GDS-1000-U Series User Manual
118
4. Press Wavetype repeatedly to
select the standard square
wave.
Wave Type
5. Press the Autoset key. The
compensation signal will
appear in the display.
Autoset
6. Press the Display key, then
Type to select the vector
waveform.
Type
Vectors
Display
7. Turn the adjustment point on the probe until
the signal edge becomes sharp.
Over
Compensation
Normal
Under
Compensation