GDS-1000-U Series User Manual
50
Using the probe compensation signal
Background
This section introduces how to use
the probe compensation signal for
general usage, in case the DUT
signal is not available or to get a
second signal for comparison. For
probe compensation details, see
page 117.
Note: The frequency accuracy and duty factor are
not guaranteed. Therefore the signal should not be
used for reference purposes.
Waveform type
Square waveform used for probe
compensation. 1k ~ 100kHz, 5% ~
95%.
Demonstration signal for showing
the effects of peak detection. See page
72 for peak detection mode details.
View the probe
compensation
waveform
1. Connect the probe between the compensation
signal output and Channel input.
VOLTS/DIV VOLTS/DIV TIME/DIV
CH 1 MATH CH 2 MENU MENU
Acquire
Display
Utility Help
Run/Stop
VARIABLE
FORCE
Autoset
Cursor
SINGLE
Hardcopy
Measure
Save/Recall
LEVEL
VERTICAL HORIZONTAL TRIGGER
CH1
CAT300V
M
W
15pF
MAX. 300Vpk
1
CH2 EXT TRIG
CAT300V
M
W
15pF
MAX. 300Vpk
1
X Y
X
10
X
1
CH1
x1
x10
2. Press the Utility key.
Utility
3. Press ProbeComp.
ProbeComp
Menu