Manual
GDS-2000 Series Training Manual
40
POSITION
9. Set the scales as follows.
Horizontal: 10ms
Vertical: CH1 200mV/div, CH2 200mV/div
10. Press the Acquire key → F5 (Mem Leng) and select long memory.
Mem Leng
12500
F 5Acquire
11. Press the trigger MENU key → F2 (Source) and select CH2.
Press F3 (Mode) and select Normal mode.
(If the waveform does not appear, increase the Trigger level)
Source
CH2
F 2
MENU
Mode
Auto
F 3
Mode
Normal
F 3
12. You can see the ripple waveform, which is an analog
representation of the ripple counter signals. Using the trigger
output from GLA-1000, the waveform is triggered at the specified
value (decimal 50 in the below example, as specified above).
Demonstration completed