254
Probe Specifications
GTP-070A-4
Applicable to: GDS-2072A & GDS-2074A
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 70MHz
Input Resistance
10Mwhen used with oscilloscopes
with 1M input.
Input Capacitance
28pF~32pF
Compensation Range
25pF~ 45pF
Max. Input Voltage
600Vpk, Derating with frequency
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 6MHz
Input Resistance
1M when used with 1M input
Input Capacitance
120pF~220pF
Max. Input Voltage
200Vpk, Derating with frequency
Safety
EN61010-031 CAT II
GTP-150A-2
Applicable to: GDS-2102A & GDS-2104A
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 150MHz
Rise Time
2.3nS
Input Resistance
10Mwhen used with oscilloscopes
with 1M input.
Input Capacitance
Approximately 17pF
Compensation Range
10 to 35pF
Max. Input Voltage
500V CAT I, 300V CAT II (DC + peak
AC) derating with frequency.
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 6MHz
Rise Time
58nS
Input Resistance
1M(Oscilloscope Input)
Input Capacitance
47pF plus oscilloscope capacitance
Max. Input Voltage
300V CAT I, 150V CAT II (DC + peak
AC) derating with frequency.
Safety
EN61010-031 CAT II