195
:DIGital:ANAlog:A<X>:DISPlay
Set
Query
Description
Turns the selected analog waveform on/off or
returns its status.
Syntax
:DIGital:ANAlog:A<x>:DISPlay { OFF | ON | ? }
Parameter
<X>
Analog waveform 1 or 2.
OFF
Turns the selected analog waveform
off.
ON
Turns the selected analog waveform
on.
Return parameter
Returns the status of the selected analog waveform
(ON, OFF).
Example
DIGital:ANAlog:A1:DISPlay on
Turns on the A1 analog waveform.
:DIGital:ANAlog:A<X>:RATio
Set
Query
Description
Sets or returns the vertical scale of the selected
analog waveform.
Syntax
:DIGital:ANAlog:A<X>:RATio {<NRf> | ? }
Parameter
<X>
Analog waveform 1 or 2.
<NRf>
Ratio (0.1, 0.2, 0.3, 0.4……1)
Return parameter
<NR2>
Returns the scale of the selected
analog waveform.
Example1
:DIGital:ANAlog:A1:RATio 0.1
Sets the analog waveform scale for A1 to 0.1.
:D<x>:MEMory
Query
Description
Returns the data in acquisition memory for the
selected digital channel as a header + raw data.