User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Utility Commands
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Go_NoGo Commands
- Replay Commands
- Rotate Commands
- DMM Commands
- APPENDX
- INDEX
GDS-200 & GDS-300 Programming Manual
124
Replay Commands
The replay function automatically logs acquisitions when the scope
is in Run mode.
:REPLAY:TOTalnum .................................................. 124
:REPLAY:CURRent ..................................................... 124
:REPLAY:TOTalnum
Description
Queries the total number of waveforms that can be
stored for the replay function.
Syntax
:REPLAY:TOTalnum?
Related
commands
REPLAY:CURRent
Return parameter
<NR1>
Returns the total number of
waveforms.
Example
:REPLAY:TOTalnum?
>3000
A maximum of 3000 waveforms can be stored for
the replay function.
:REPLAY:CURRent
Description
Sets or queries the current waveform that is
displayed when using the replay function. Note:
the scope may need to be in Stop mode to use this
function.
Syntax
REPLAY:CURRent { FORward | BACKward | <NRf> | ? }
Related
commands
REPLAY:TOTalnum?
Parameter/Return
parameter
FORward
Moves to the next waveform in
memory
Query
Set
Query