User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
131
MANU:ACW:FREQuency
Set
Query
Description
Sets or returns the ACW test frequency in Hz. The
test must first be in ACW mode before this
command can be used.
Syntax
Query Syntax
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/
Return parameter
<50>
50 Hz
<60>
60 Hz
Example
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
MANU:ACW:REF
Set
Query
Description
Sets or returns the ACW reference value in mA.
The test must first be in ACW mode before this
command can be used.
The ACW reference value must be less than the HI
SET value.
The ACW reference value must use the same range
as the HI SET value.
Syntax
Query Syntax
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX)
Example
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.