User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
140
MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The
test must first be in GB mode before this command
can be used.
The GB reference value must be less than the HI
SET value.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:REF 100
Sets the GB reference to 100 mΩ.
MANU:GB:ZEROCHECK
Set
Query
Description
Performs the zero check function. The test must
first be in GB mode and in the Ready Status before
this command can be used.
See page 71 for details on the ZERO function.
Syntax
Query Syntax
MANU:GB:ZEROCHECK {ON|OFF}
MANU:GB:ZEROCHECK?
Parameter/
Return parameter
<ON>
Zero function is active.
<OFF>
Zero function is not active.
Example
MANU:GB:ZEROCHECK OFF
Activates the ZERO function.