User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
142
Parameter/
Return parameter
CONT
Sets/returns the fail mode as continue.
HOLD
Sets/returns the fail mode as hold.
STOP
Sets/returns the fail mode as stop.
Example
MANU:UTIL:FAIL CONT
Sets the fail mode to CONT (continue).
MANU:UTILity:MAXHold
Set
Query
Description
Sets or returns the MAX HOLD setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:MAXHold {ON|OFF}
MANU:UTILity:MAXHold?
Parameter/
Return parameter
OFF
Turns MAX HOLD off.
ON
Turns MAX HOLD on.
Example
MANU:UTIL:MAXH ON
Turns MAX HOLD on.
MANU:UTILity:GROUNDMODE
Set
Query
Description
Sets or returns the Grounding mode of the current
test.
The Ground Mode setting cannot be turned on
with the IR and GB function.
Syntax
Query Syntax
MANU:UTILity:GROUNDMODE {ON|OFF}
MANU:UTILity:GROUNDMODE?
Parameter/
Return parameter
OFF
Turns ground mode off.
ON
Turns ground mode on.
Example
MANU:UTIL:GROUNDMODE ON
Turns GROUND MODE on.