User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
47
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set the
TIMER value.
ACW
DCW
IR
GB
000.5s~999.9s
000.5s~999.9s
001.0s~999.9s
000.5s~999.9s
Note
With the ACW test function, when the test current
is between 30mA and 40mA (GPT-98XX) or 80mA
and 100mA (GPT-99XX), the ramp time + test time
cannot exceed 240 seconds. At this current level,
the tester also needs to pause after a test for a
time equal to or greater than the output time. See
the specifications on page 161 for details.
Special Manual
Mode
When in special MANU test mode (page 71) the
Timer can be turned off when using the DCW
or ACW test function.
Hold the TIMER soft-key for 3
seconds to turn the timer off.
I M ET R
Note
The timer can only be turned off under special
MANU test mode, however there is a limitation:
The timer cannot be turned off (limited to 240s) if
the test current is between 30mA and 40mA (GPT-
98XX) or 80mA and 100mA (GPT-99XX) in ACW
mode.
The discharge time and initial test time cannot be
edited.