User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
93
(excluding skipped tests).
A FAIL result for a single step will result in
FAIL for the whole automatic test.
A STOP. No step can be stopped for a
PASS/FAIL judgment to be made. In other
words, if a test is stopped, it is judged as neither
PASS nor FAIL.
No step can contain an ERROR or ILOCK
message.
E R R
P A S
O
S
P A S
S K I
S
P
E MA U OT N A_ 1A U T O = 0 0 1 - 0 0
# 0 1 : # 0 2 : # 0 3 : # 0 4 : P A S
# 0 5 : # 0 6 : # 0 7 : # 0 8 : S T O
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
S
P
I L O
F A I
C
L
ILOCK messageERROR message
R K
ERROR: Indicates that V, I or R is not correct.
This usually occurs if the testing leads are not
properly connected.
ILOCK: Indicates that the interlock key is
disconnected (if configured to be used).
PASS Judgment
When all the tests have been
judged as PASS, the PASS
indicator will be lit green and the
buzzer will sound.
PASS
P A S
P A S
S
S
P A S
P A S
S
S
E MA U OT N A_ *A U T O = 0 0 1 - * *
# 0 1 : # 0 2 : # 0 3 : # 0 4 : P A S
# 0 5 : # 0 6 : # 0 7 : # 0 8 : P A S
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
S
S
P A S
P A S
S
S
Note
The Pass Sound setting must to set to ON for the
buzzer to sound (page 97).
FAIL Judgment
When any of the tests have been
judged as FAIL, the FAIL indicator
will be lit red and the buzzer will
sound.
FAIL