Running a Manual Test
Background
A test can be run when the tester is in READY
status.
Note
The tester cannot start to run a test under the
following conditions:
A protection setting has been tripped; when a
protection setting has been tripped the
corresponding error message is displayed on
the screen.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O
port.
The STOP signal has been received remotely.
If Double Action is ON, ensure the START
button is pressed immediately after the STOP
button (<0.5s).
Note
When a test is running the voltage output cannot
be changed, unless the test is under the special
manual mode.
Steps
1. Ensure a MANU test is selected and the tester is
in VIEW status.