User Manual

GSB-01/GSB-02 Series User Manual
50
FAIL Judgment
3. When the test is judged as FAIL,
FAIL will be displayed on the
safety tester, the FAIL indicator
will be lit red, the channel
indicator on the scanner will be lit
red and the buzzer will sound.
As soon as a test is judged FAIL,
power is cut from the terminals.
FAIL
Safety tester
example
I R G B A CW D C W
F R E Q = 0 H z 6
0
100
k V
IAF
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
2M A UN 0 0_2M A N U = * * * - 0 0
L
m A
01 37
Scanner display
example
Multiplex Scanner Box
GSB-01
H I
LO
H I
LO
H I
LO
H I
LO
H I
LO
H I
LO
H I
LO
H I
LO
CHANNEL 2 3 4 5 6 7 8
PASS
FAIL
1
Fail
4. The FAIL judgment will be held on
the safety tester display and the
scanner(s) until the STOP button is
pressed. Pressing the STOP button
twice will return the tester to the
READY status.
STOP
5. The READY indicator will be lit
blue in the READY status.
READY
I R G B
m A
A CW D CW
E
F R E Q = 0 H z 6
0
100
k V
AER
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
D
MM A UN N A_2M A N U = * * * - 0 0
READY status
Y
Note