User Manual
OPERATION
51
Automatic Tests
This section describes how to create and run automatic tests for the
GSB-01/GSB-02 scanner boxes.
An automatic test comprises of up 16 MANU tests that are run
sequentially. In order to run an automatic test using the GSB-
01/GSB-02 scanner boxes, the scanner outputs for each manual test
must first be configured, and then each manual test in the automatic
test can be stepped through. In other words, the channel outputs
are directed by the manual tests that comprise the automatic test.
It is assumed that you have read and understood how to create an
automatic test on the GPT-9000 safety tester. If you are not sure
how to create and run an automatic test on the safety tester, please
see the GPT-9000 user manual.
Before operating the GPT-9000 please read the safety precautions as
outlined in the Set Up chapter on page 18.
Create an Automatic Test
Background
It is assumed that you already know how to
create an automatic test. This section will only
give the essential instructions for creating an
automatic test. For more comprehensive details,
please see the GTP-9000 user manual.
Steps
1. If the tester is in MANU mode,
press and hold the MANU/AUTO
key for three seconds. This will put
the tester into Auto mode.
The tester can only switch between
AUTO and MANU mode when in
the VIEW status.
MANU/AUTO