236
5
Load On
Load On Input.
Load on = Active low, 0-1V
Load off = Active high, 4-5V.
(Using the + load polarity in
conjunction with a 10k pull up
resistor)
6
+15V
Internal power output. Max
50mA.
Channel Control
Interface
Constraints
Mode/Range
Mode and Range configuration
is only selected via the front
panel.
Configure Frame Link Interface
Connection
Connect the Frame link cable
(MIL 20 pin connector) to the
rear panel port: 20-pin male
connector.
Pin assignment
(Frame link
connector 1)
Pin1
A Recall Preset memory 0
(All channels)
Pin2
B Recall Preset memory 1
(All channels)
Pin3
C Recall Preset memory 2
(All channels)
Pin4
TRIG_OUT Trigger out
Pin5
MEM_0 Recall Setup memory 1
(All channels)
Pin6
MEM_1 Recall Setup memory 2
(All channels)