Current 0~7A 0~70A 0~1A 0~10A
Current
Resolution
1mA 10mA 0.2mA 2mA
Current Accuracy
±0.4% F.S. ±0.4% F.S.
MEASUREMENT
VOLTAGE READBACK
Range 0~16V 0~80V 0~125V 0~500V
Resolution 0.5mV 2.5mV 4mV 16mV
Accuracy ±(0.05%set + 0.05% F.S.) ±(0.05%set + 0.05% F.S.)
CURRENT READBACK
Range 0~7A 0~70A 0~1A 0~10A
Resolution 0.175mA 1.75mA 0.032mA 0.32mA
Accuracy ±(0.1%set + 0.1% F.S.) ±(0.1%set + 0.1% F.S.)
PROTECTIVE
Over Power Protection
Range 1.75~357W 1.75~357W
Resolution 1.75W 1.75W
Accuracy ±(2%set + 0.25%F.S) ±(2%set + 0.25%F.S)
Over Current Protection
Range 0.875~71.4A 0.125~10.2A
Resolution 0.175A 0.025A
Accuracy ±(2%set + 0.25%F.S) ±(2%set + 0.25%F.S)
Over Voltage Protection
Range 1~81.6V 2.5~500V
Resolution 0.2V 1.25V
Accuracy ±(2%set + 0.25%F.S) ±(2%set + 0.25%F.S)
Over
Temperature
Protection
ؒ85Ԩ ؒ85Ԩ
Rated Power Protection
Value 385W 385W
Accuracy ±2%set ±2%set ±2%set
GENERAL
SHORT CIRCUIT
Current(CC)
ؒ7.7/7A ؒ77/70A ؒ1.1/1A ؒ11/10A
Voltage(CV) 0V 0V 0V 0V
Resistance(CR)
ؒ1.25 ؒ0.025 ؒ50 ؒ1.25
INPUT
RESISTANCE
(LOAD OFF)
800K(Typical) 800K(Typical)