Go/NoGo Connection
Background
The Go/NoGo port is a 15 socket port. Each
channel has a dedicated line for a Go/NoGo
output. The ports are open-collector with active
low (1.1V) indicating a pass and active high (30V)
as fail (an alarm). The Go/NoGo terminal is a DB-
15 female.
For more details on the Go/NoGo interface see
page 239.