User manual

REMOTE CONTROL
161
2. Connect a GPIB cable from a GPIB controller to
the GPIB port on the PEL-3000.
3. Turn the PEL-3000 on.
4. Press
Shift
+
Utility
Help
> Interface[F3] and set
the Interface setting to GPIB.
5. Set the GPIB address.
GPIB address
0~30
GPIB constraints
Maximum 15 devices altogether, 20m cable length,
2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
Pin Assignment
112
1324
Pin
Signal
Pin
Signal
1~4
Data I/O 1~4
13~16
Data I/O 5~8
5
EOI
17
REN
6
DAV
18
Ground (DAV)
7
NRFD
19
Ground (NRFD)
8
NDAC
20
Ground (NDAC)
9
IFC
21
Ground (IFC)
10
SRQ
22
Ground (SRQ)
11
ATN
23
Ground (ATN)
12
SHIELD Ground
24
Single GND