User manual

COMMNAND OVERVIEW
85
Related
Commands
:RESistance:RISE
Parameter
<NRf>
Falling current slew rate
[ mA/us]
MINIMUM or MIN
Minimum slew rate
MAXIMUM or MAX
Maximum slew rate
Return parameter
<NR2>
Returns the falling current
slew rate in mA/us.
Example
:RESistance:FALL MIN
Sets the falling slew rate to the minimum.
Query example
:RESistance:FALL?
>50.000
Returns the falling slew rate as 50mA/us.
:RESistance:T1
Set
Query
Description
Sets the T1 timer settings for the CR dynamic
mode.
Syntax
:RESistance:T1 {<NRf>[s] | MINimum | MAXimum}
Query Syntax
:RESistance:T1?
Related
Commands
:RESistance:T2
Parameter
<NRf>[s]
T1 timer setting in
seconds
MINIMUM or MIN
Minimum time
MAXIMUM or MAX
Maximum time
Return parameter
<NR2>
Returns the T1 timer time
in seconds.
Example
:CURRent:T1 0.200
:CURRent:T1 0.200s
:CURRent:T1 200 ms
Sets the T1 timer time to 200ms.