User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMNAND OVERVIEW
- Command Syntax
- Command List
- Common Commands
- Trigger Commands
- Input State Commands
- Measurement Commands
- Configure Subsystem Commands
- Parallel Commands
- Step Resolution Commands
- External Control Commands
- Mode Subsystem Commands
- Current Subsystem Commands
- Resistance Subsystem Commands
- Voltage Subsystem Commands
- Power Subsystem Commands
- Function Commands
- Utility Commands
- File Commands
- SCPI Register Commands
- Status Registers
- Error Messages
PEL-3000 Programming Manual
96
Example
:POWer:T2 0.200
Sets the T2 timer time to 200ms.
Query example
:POWer:T2?
>0.200
Returns the T2 timer time as 200ms.
:POWer:FREQuency
Set
Query
Description
Sets the CP dynamic mode switching frequency.
This command along with the “:POWer:Duty”
command is used when the dynamic time setting
is set to “Freq./Duty”.
This command is only applicable for dynamic
mode.
Syntax
:POWer:FREQuency {<NRf> | MINimum | MAXimum}
Query Syntax
:POWer:FREQuency?
Related
Commands
:POWer:DUTY
Parameter
<NRf>
Sets the switching
frequency in hertz.
MINIMUM or MIN
Minimum frequency
MAXIMUM or MAX
Maximum frequency
Return parameter
<NR2>
Returns the frequency in
hertz.
Example
:POWer:FREQuency 60
Sets the frequency to 60Hz.
Query example
:POWer:FREQuency?
>60