Manual

REMOTE CONTROL
101
Command
:SENSe[1]:PCURrent:SYNChronize[:STATe]?
Function
Queries the pulse current measurement triggering
option.
Example
:SENSe:PCURrent:SYNChronize?
Returns the pulse current trigger option.
Command
:SENSe[1]:PCURrent:SYNChronize:DELay <NRf>
Function
Sets the trigger delay time.
Description
<NRf>
0~0.1 or 0~5 (Pulse current digitization)
Example
:SENSe:PCURrent:SYNChronize:DELay 0.05
Sets the trigger delay time to 0.05 seconds.
Command
:SENSe[1]:PCURrent:SYNChronize:DELay?
Function
Queries the trigger delay time.
Example
:SENSe:PCURrent:SYNChronize:DELay
Returns the trigger delay time.
Command
:SENSe[1]:PCURrent:SYNChronize:TLEVel<NRf>
Function
Sets the trigger level.
Description
<NRf>
0.000-5.000
Example
:SENSe:PCURrent:SYNChronize:TLEVel 1
Sets the trigger level to 1.000A.