Manual

REMOTE CONTROL
93
Source Commands
Command
:[SOURce]:CURRent[:LIMit][:VALue] <NRf>
Description
Sets the current level.
NRf
0.0000-5.0000
Example
:SOURce:CURRent 1.0005
Sets the current level to 1.0005A.
Command
:[SOURce]:CURRent[:LIMit][:VALue]?
Description
Queries the current limit level.
Example
:SOURce:CURRent?
Returns the current limit level.
Command
:[SOURce]:CURRent[:LIMit]:TYPE <name>
Function
Sets the current limit mode.
Description
<name>
LIMit: General limiting mode
TRIP: Output shutdown mode
LIMRELAY|LIMITRELAYGeneral
limiting mode and external relay output
control mode.
TRIPRELAY: Ouput shutdown mode
and external relay output control mode.
Example
:SOURce:CURRent:TYPE LIMITRELAY
Sets the current limit mode to LIMITRELAY.