Manual

PPH-1503 User Manual
96
Command
:SENSe[1]:NPLCycles?
Function
Returns the number of power line cycles used for
the integration rate.
Example
:SENSe:NPLCycles?
Returns the number of PLCs.
Command
:SENSe[1]:AVERage <NRf>
Function
Sets the averaging number for the voltage, current
and DVM measurements.
Description
<NRf>
1-10
Example
:SENSe:AVERage 3
Sets the averaging number to 3.
Command
:SENSe[1]:AVERage?
Function
Returns the average number.
Examle
:SENSe:AVERage?
Returns the average number.
Command
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer] <n>
Description
Sets the current measurement range.
Description
<n>
MINlow range
MAXhigh range
Example
:SENSe:CURRent:RANGe MIN
Sets the current range to low.
Command
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer]?