Connection- alt.
shielding
If the wire shield needs to be grounded at the
voltage source (EXT-V), then the shield cannot
also be grounded at the negative (-) terminal
output of the PSU power supply. This would
short the output.
PSUEXT-V
Analog
connector
23
21
Output
Terminal
2 core shielded
wire or twisted
pair
Pin23 → EXT-V (-)
Pin21 → EXT-V (+)
Wire shield → EXT-V ground (GND)
Steps
1. Connect the external voltage according to the
connection diagrams above.
2. Set the F-91 power on
configuration setting to 1 (CC
control Ext voltage).
Page 111
Be sure to cycle the power after the power
on configuration has been set.
3. Press the Function key and confirm
the new configuration settings (F-
91=1).
Function
4. Press the Output key. The
current can now be
controlled with the
External voltage.
Output