Isolated External Voltage Control of Current Output
Background
Voltage control of the current output uses the
isolated voltage option (PSU-ISO-V). A voltage
of 0~5V or 0~10V is used to control the full
scale current of the instrument, where:
For 0~5V: Output current = full scale current ×
(external voltage/5)
For 0~10V: Output current = full scale current ×
(external voltage/10)
Connection
When connecting the external voltage source to
the isolated voltage option, use shielded or
twisted paired wiring.
PSUEXT-V
Isolated
Voltage
Option
4
3
Output
Terminal
2 core shielded
wire or twisted
pair
1or8
Pin4(GND_ISO) → EXT-V (-)
Pin3(+IPROG_ISO) → EXT-V (+)
EXT-V ground (Shield) Wire shield