Source Commands
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] 54
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] . 55
[SOURce:]CURRent:PROTection[:LEVel] ........................ 55
[SOURce:]CURRent:PROTection:STATe ......................... 56
[SOURce:]CURRent:SLEW:RISing ..................................... 56
[SOURce:]CURRent:SLEW:FALLing ................................ 56
[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]57
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] 57
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] . 58
[SOURce:]VOLTage:PROTection[:LEVel] ....................... 58
[SOURce:]VOLTage:SLEW:RISing .................................... 59
[SOURce:]VOLTage:SLEW:FALLing ................................ 59
[SOURce:]CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NRf>|MIN|MAX}
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
[MIN|MAX]
<NRf>
0~105% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
SOUR:CURR:LEV:IMM:AMPL? MAX
37.800
Returns the maximum possible current level in amps.